Chromatograph

(redirected from chromatographic)
Also found in: Dictionary, Thesaurus, Medical.

chromatograph

[krō′mad·ə‚graf]
(analytical chemistry)
To employ chromatography to separate substances.

Chromatograph

 

a device for the chromatographic separation and analysis of mixtures. The principal parts of a chromatograph include a system for injecting the sample mixture, the chromatographic column, a detection device (detector), and recording and thermostatic control systems; in addition, preparative chromatographs (including production installations) also feature disassembly fittings and receivers for the separated components.

Chromatographs are classified according to the state of aggregation of the mobile phase as gas and liquid chromatographs. Development chromatography is the method used in the vast majority of cases.

In a gas chromatograph (see Figure 1) the carrier gas is fed continuously from a cylinder through flow and pressure regulators at constant or variable velocity into the chromatographic column (a tube 2–5 mm in diameter and 1–10 m long); the column is filled with a sorbent and placed in a thermostat that makes it possible to maintain a specified temperature (up to 500°C).

Injection of a gas sample (1–50 cm3) or liquid sample (a few microliters) is carried out either manually (with a gas injector or hypodermic needle), or automatically (by means of batchers). The original multicomponent mixture separates in the chromatographic column into a series of two-component mixtures consisting of the carrier gas and one of the components being analyzed. The mixtures enter the detector in a specific sequence that depends on the different adsorption capacities of the components. As a result of the processes that occur in the detector, such as changes in thermal conductivity and ionization flow, a change in the concentration of the effluent components is recorded. Converted into an electrical signal, these processes are recorded as the output curve.

The most commonly used detectors in gas chromatographs are the thermal conductivity and ionization types. A typical thermal conductivity detector (katharometer) has a bridge circuit connected to two cells for measuring thermal conductivity. The flows of pure carrier gas and of the two-component mixture pass through the cells. The thermal conductivity of the two-component mixture differs from that of the pure carrier gas. When the mixture passes through the sensing element of the detector—a heated coil with a resistance of 10–80 ohms—the temperature and resistance of the coil change as a function of the concentration of the component. Such detectors make it possible to determine concentrations in the range 10–1–10–2 percent.

Figure 1. Diagram of a gas chromatograph: (1) cylinder containing an inert gas, (2) device for injecting the sample into the chromatographic column, (3) chromatographic column, (4) thermostat, (5) detector, (6) signal converter, (7) recorder

The principal part of ionization detectors is the ionization chamber, where the molecules that enter the chamber with the carrier gas from the chromatographic column are ionized. Ionization of the substances being studied is carried out in a hydrogen flame or by means of metastable argon or helium atoms, slow electrons, and so forth. Under the applied voltage, the ions move in the ionization chamber, creating an electric current. Ionization detectors make it possible to determine concentrations of substances in the range 10–4–10–7 percent.

Thermal conductivity and ionization detectors may be characterized with respect to their sensitivity (the minimum determinable concentration of a substance), selectivity (the ability selectively to determine individual components in a mixture), and the direct dependence of signal on concentration.

In a liquid chromatograph the detecting device may be a flow refractometer connected via a differential circuit or an ultraviolet absorption detector. The mobile phase solvent is fed by means of pulse-free systems (at pressures up to 50 meganewtons per m2, or 500 kg-force per cm2), and sample injection is accomplished with a hypodermic needle or switching valve. The length of the chromatographic column in a liquid chromatograph does not exceed 1 m. On the whole, the detectors of liquid chromatographs have much lower sensitivity (approximately two orders of magnitude) than those used in gas chromatographs. For precise measurement of the concentrations of substances, the detectors are calibrated with respect to mixtures of known composition.

The attainable speed and precision of analysis in a chromatograph are determined largely by the proper selection of the operating conditions of the detector and of the experiment conditions, such as the type of sorbent, temperature, carrier gas rate, and length of the chromatographic column. A time-programmed change in the temperature of the chromatographic column or flow of the carrier gas is used to accelerate analysis.

REFERENCES

Pribory dlia khromatografii. Moscow, 1973.
Brazhnikov, V. V. Differentsial’nye detektory dlia gazovoi khromatografii. Moscow, 1974.

K. I. SAKODYNSKH

References in periodicals archive ?
Additionally, the ACQUITY QDa Detector extends the limits of sample detection for a chromatographic separation by quantifying compounds having no UV response or that are present at levels not detectable and quantifiable by optical detection.
Over the last decade, demand for advanced and innovative chromatographic technologies has increased considerably due to higher adoption of ion exchange and flash chromatography.
These data demonstrate that the addition of a chromatographic surface in our method leads to the binding of more proteomic biomarkers, thus providing better differentiation between pathological entities.
The only pertinent weakness, besides the expense of instrumentation and the labeled standards, is the inability to qualitatively distinguish among certain isomers that have similar chromatographic properties and mass spectral fragmentation patterns (i.
Columns of various sizes were packed with the ground MIP material, either as a chromatographic stationary phase or as a solid phase adsorbent.
Chapter 20 Discontinuous Moving Bed Chromatographic Reactors
Twenty-two American academics and researchers contribute 22 chapters providing guidelines for toxicologists, pathologists, and laboratory scientists regarding the application of chromatographic techniques for therapeutic drug monitoring of both classical and newer drugs.
The first would involve starting with a method developed for determining drug product impurities, then identifying operating parameters that can expand the chromatographic range to provide additional separation capability for extractable compounds.
As a result, all endogenous amino compounds produce fluorescent derivatives, which produce very large solvent peaks, decreasing the sensitivity and increasing the length of the chromatographic run times.
a supplier of high-capacity chromatographic media for Millipore's Prosep-A High Capacity Chromatographic Media.
The methods currently used yield only averages of the mass distribution; chromatographic methods produce the mass distribution, if calibrated properly with polymer mass standards.