(analytical chemistry)
To employ chromatography to separate substances.



a device for the chromatographic separation and analysis of mixtures. The principal parts of a chromatograph include a system for injecting the sample mixture, the chromatographic column, a detection device (detector), and recording and thermostatic control systems; in addition, preparative chromatographs (including production installations) also feature disassembly fittings and receivers for the separated components.

Chromatographs are classified according to the state of aggregation of the mobile phase as gas and liquid chromatographs. Development chromatography is the method used in the vast majority of cases.

In a gas chromatograph (see Figure 1) the carrier gas is fed continuously from a cylinder through flow and pressure regulators at constant or variable velocity into the chromatographic column (a tube 2–5 mm in diameter and 1–10 m long); the column is filled with a sorbent and placed in a thermostat that makes it possible to maintain a specified temperature (up to 500°C).

Injection of a gas sample (1–50 cm3) or liquid sample (a few microliters) is carried out either manually (with a gas injector or hypodermic needle), or automatically (by means of batchers). The original multicomponent mixture separates in the chromatographic column into a series of two-component mixtures consisting of the carrier gas and one of the components being analyzed. The mixtures enter the detector in a specific sequence that depends on the different adsorption capacities of the components. As a result of the processes that occur in the detector, such as changes in thermal conductivity and ionization flow, a change in the concentration of the effluent components is recorded. Converted into an electrical signal, these processes are recorded as the output curve.

The most commonly used detectors in gas chromatographs are the thermal conductivity and ionization types. A typical thermal conductivity detector (katharometer) has a bridge circuit connected to two cells for measuring thermal conductivity. The flows of pure carrier gas and of the two-component mixture pass through the cells. The thermal conductivity of the two-component mixture differs from that of the pure carrier gas. When the mixture passes through the sensing element of the detector—a heated coil with a resistance of 10–80 ohms—the temperature and resistance of the coil change as a function of the concentration of the component. Such detectors make it possible to determine concentrations in the range 10–1–10–2 percent.

Figure 1. Diagram of a gas chromatograph: (1) cylinder containing an inert gas, (2) device for injecting the sample into the chromatographic column, (3) chromatographic column, (4) thermostat, (5) detector, (6) signal converter, (7) recorder

The principal part of ionization detectors is the ionization chamber, where the molecules that enter the chamber with the carrier gas from the chromatographic column are ionized. Ionization of the substances being studied is carried out in a hydrogen flame or by means of metastable argon or helium atoms, slow electrons, and so forth. Under the applied voltage, the ions move in the ionization chamber, creating an electric current. Ionization detectors make it possible to determine concentrations of substances in the range 10–4–10–7 percent.

Thermal conductivity and ionization detectors may be characterized with respect to their sensitivity (the minimum determinable concentration of a substance), selectivity (the ability selectively to determine individual components in a mixture), and the direct dependence of signal on concentration.

In a liquid chromatograph the detecting device may be a flow refractometer connected via a differential circuit or an ultraviolet absorption detector. The mobile phase solvent is fed by means of pulse-free systems (at pressures up to 50 meganewtons per m2, or 500 kg-force per cm2), and sample injection is accomplished with a hypodermic needle or switching valve. The length of the chromatographic column in a liquid chromatograph does not exceed 1 m. On the whole, the detectors of liquid chromatographs have much lower sensitivity (approximately two orders of magnitude) than those used in gas chromatographs. For precise measurement of the concentrations of substances, the detectors are calibrated with respect to mixtures of known composition.

The attainable speed and precision of analysis in a chromatograph are determined largely by the proper selection of the operating conditions of the detector and of the experiment conditions, such as the type of sorbent, temperature, carrier gas rate, and length of the chromatographic column. A time-programmed change in the temperature of the chromatographic column or flow of the carrier gas is used to accelerate analysis.


Pribory dlia khromatografii. Moscow, 1973.
Brazhnikov, V. V. Differentsial’nye detektory dlia gazovoi khromatografii. Moscow, 1974.


References in periodicals archive ?
A gas chromatograph that is capable of calculating the HCDP at any pressure enables a very useful operational diagnostic that can be used to avoid two-phase flow, and therefore incorrect measurement.
Rosemount Analytical's GCX Gas Chromatograph Transmitter has obtained FOUNDATION(TM) fieldbus certification and is now authorized to bear the FOUNDATION fieldbus registration checkmark after completing the rigorous Fieldbus Foundation testing.
To manage data, quality control departments in the pharmaceutical industry increasingly need to integrate the data from multiple chromatographs used in their laboratories.
Package 1 - Supply of liquid chromatograph mass spectrophotometer with the necessary equipment, package 2 - supply of gas chromatograph with quadrupole analyzer with the necessary equipment package 3 - dual-channel supply gas chromatographs together with a computer program, computer and additional equipment necessary for work.
During the term of the agreement, Siemens (NYSE: SI)will supply gas chromatographs together with related systems such as analysis cabinets and shelters, much of which will be manufactured by Siemens in the United States.
Many gas chromatographs perform a complete analysis for all of the major components expected in the sample, such as the natural gas "Btu" measurement.
Keywords in this release: process analytical instruments, software and digital communications, value engineering, data management firms, bio-technology, chemicals, gas refining industries, chromatographs, spectrometer, gas analyzers, liquid analyzers, research, information, market, trends, technology, service, forecast, market share
It is the latest in Emerson's line of transmitter-style gas chromatographs that also includes the Model 500 and Model 700.
ABB Totalflow has released its NGC 8200 series of online natural gas chromatographs.
NYSE:A) today announced a joint initiative to create NI-certified LabVIEW instrument drivers for control of Agilent gas and liquid chromatographs.
Tenders are invited for Spectrophotometers, Chromatographs, Analyzers and Related Specialized Laboratory Equipment/Accessorie