Luminance Meter

Luminance Meter

 

a photometer that is used to measure luminance, or photometric brightness. Two optical diagrams of luminance meters with physical radiation detectors are shown in Figure 1,c and d in PHOTOMETER.

In a luminance meter built according to the optical design in Figure 1,c, the image of the luminous body in the light source (S) is formed in the plane of the diaphragm D, which limits the size of the part of the body being measured. When the objective is moved, the sensitivity of such a luminance meter is kept constant by means of an aperture stop Da, which is fixed with respect to D.

In a simpler luminance meter, which is built according to the optical design in Figure 1,d, the bundle of rays being measured is limited by means of both a calibrated diaphragm Dc and the entrance pupil (see) of the radiation detector. The calibrated diaphragm Dc is placed either near the luminous body or, in the photometry of large objects, at some distance from the luminous body. The angle of the field of view of such a luminance meter may be varied discretely by superimposing a converging lens with a focal length of l0 on the calibrated diaphragm. In this way, such a luminance meter is converted into a photometer corresponding to the optical design in Figure 1,c.

The human eye or an animal’s eye, either of which directly perceives brightness, is the simplest type of visual luminance meter. An equivalent optical diagram of such a luminance meter is shown in Figure 1,c.

Several types of luminance meters are manufactured by industry. They include photometers for measuring the luminance of continuous or pulsed light sources, visual photometers for measuring the equality of brightness, optical pyrometers, and exposure meters, which are used in photography. An exposure meter may be built into a camera or may be a separate instrument.

REFERENCES

Gorokhovskii, Iu. N., B. N. Orlov, and A. M. Freivert. “Iarkomery s peremennym uglom vospriiatiia dlia eksponometricheskikh tselei.” Zhurnal nauchnoi i prikladnoi fotografii i kinematografii, 1974, vol. 19, no. 5.
See also references under PHOTOMETRY.

A. S. DOINIKOV

luminance meter, brightness meter

A visual instrument or a photoelectric instrument used to measure luminance.
Mentioned in ?
References in periodicals archive ?
Limited Tenders are invited for Calibration Of Luminance Meter At Air Force Station Hindan Ghaziabad
An exterior luminance meter has been installed to measure the amount of natural daylight, whilst a lighting flux regulator adjusts the levels inside the tunnel.
The effective luminance of the stimulus was calibrated with a Konica Minolta LS-100 Luminance Meter (Konica Minolta Holdings, Inc.
These photos are combined in software such as Photosphere, and calibrated with a luminance meter against a uniformly lit reflectance standard.
The luminance was measured by a luminance meter (Konica Minolta LS110) with the objective directed towards the ring.
We, however, used a Minolta LS100 luminance meter to obtain control luminances in the higher and lower extremes of the dynamic range to calibrate the scenes.
The optical characteristics of the replicated LGPs were investigated by measuring the spatial luminance of the LGPs by means of a luminance meter (BM-7, Topcon) as shown in Fig.
8 After initially matching the image settings using subjective impressions in our department, the standardized look was further defined using luminance meter measurements obtained directly from the soft copy view station display (Figure 3).
One simple system uses an SMU connected to the OLED and a picoammeter to read the current from a photodiode or a luminance meter to measure the light output, both connected to a PC.
The luminance of the symbols against a dark background, measured using an LMT L1009 digital luminance meter, was 11.
This technique is also used more and more by lighting researchers in building applications because it offers the opportunity to rapidly capture a larger field of real world luminances than a spot luminance meter [Anaokar and Moeck 2005; Inanici 2006; Jacobs 2007].
The configuration shown includes a luminance meter used to confirm the accuracy of the results-this would not normally be required in a field application.