Reference Electrode


Also found in: Dictionary, Thesaurus, Medical, Wikipedia.
Related to Reference Electrode: working electrode, Indicator electrode

reference electrode

[′ref·rəns i′lek‚trōd]
(physical chemistry)
A nonpolarizable electrode that generates highly reproducible potentials; used for pH measurements and polarographic analyses; examples are the calomel electrode, silver-silver chloride electrode, and mercury pool.

Electrode, Reference

 

a galvanic electrode used for measuring electrode potentials. In most cases, a measurement is made of the potential difference between the electrode being tested and a selected reference electrode with a known potential relative to the potential of a normal hydrogen electrode, which by convention is taken as zero (in a stricter sense, zero is accepted as the potential of a standard hydrogen electrode, which differs from a normal hydrogen electrode in that the activity of the H+ ions, rather than the concentration, is equal to unity). The measured difference is taken as the potential of the electrode being tested, and the specific reference electrode against which the potential was measured is specified.

Electrodes selected for reference purposes have potentials that are stable and reproducible. They may be classified according to the nature of the electrochemical reactions that take place in them. The reactions must be highly reversible in order to avoid a change of potential when a small current flows through the reference electrode.

The reference electrodes that are most often used include calomel electrodes (HgǀHg2Cl2ǀKCl or HCl), silver chloride electrodes (AgǀAgClǀKCl or HCl), mercuric sulphate electrodes (HgǀHgSO4ǀH2SO4), mercury oxide electrodes (HgǀHgOǀKOH), and quinhydrone electrodes (Ptǀhydro-quinone, quinone ǀ HCI).

The potentials of reference electrodes depend on the concentration of potential-determining ions. For example, in calomel reference electrodes the potential depends on the concentration of Cl ions. The potentials of 0.1 N type, 1 N type, and saturated calomel reference electrodes at 25°C are equal, respectively, to 333, 280, and 241 millivolts (mV) relative to a normal hydrogen electrode. The variation in the potentials φ of reference electrodes with temperature t in °C is described by temperature coefficients, which are different for the various electrode types. In the case of a 1 N type calomel reference electrode, for example, φ = + 280 – 0.24 (t – 25) mV relative to a normal hydrogen electrode (NHE) at the same temperature (by definition φNHE = 0 for all temperatures).

The selection of a reference electrode depends on the measurement conditions. In nonaqueous media it is possible to use an aqueous reference electrode, but in this case the diffusion potentials at the boundary between the aqueous and nonaqueous solutions must be taken into account. Metal electrodes are used in fused electrolytes; they have potentials that do not vary with time in the particular electrolyte.

REFERENCES

Antropov, L. I. Teoreticheskaia elektrokhimiia, 3rd ed. Moscow, 1975.
Reference electrodes. Edited by D. G. Ives and G. J. Janz. New York-London, 1961.
Butler, J. “Elektrody sravneniia v aprotonnykh organicheskikh rastvoriteliakh.” In Elektrokhimiia metallov v nevodnykh rastvorakh. Moscow, 1974. (Translated from English.)

G. M. FLORIANOVICH

References in periodicals archive ?
Using an external reference electrode, the chip reliably digitizes the pH level (Figure 7).
An improved estimated [E.sub.cell(ORP)] is observed with the proposed approach of a divided, Open-circuit potentiometric cell, via Hg/[Hg.sub.2]S[O.sub.4] reference electrodes and graphite rod working electrodes.
The amperometric three-electrode glucose sensor, including working, counter, and reference electrodes, is based on the glucose oxidase method using hydrogen peroxide detection with electrochemical recycling of the oxygen, thus making the sensor essentially free from interference from the oxygen content of the sample.
The saturated calomel electrode was used as the reference electrode and the platinum electrode as the indicator electrode.
[Z.sub.e1] and [Z.sub.e2] stand for the interface impedance between the skin and reference electrode at the points of q and r respectively.
An electrochemical biosensor employs three bare ring electrodes that circulate the microchannel: a reference electrode, a working electrode (cathode or anode), and a counter electrode.
The EC Cell Kit includes the EC cell with volume reducing insert, a variety of sample mounts, a specialized all-PEEK cantilever holder and a fully-sealed liquid-filled Ag-AgCl reference electrode with PEEK barrel.
The potential of corrosion can be directly measured in report to a reference electrode that is characterised by a potential half-cell very stable.
The new reference electrode construction includes an ion barrier to protect and stabilise the reference potential in harsh conditions.
A reference electrode was positioned at some distance (approximately 25 cm) from the animal.
Saturated calomel electrode (SCE) was used as reference electrode. CV and EIS measurements were performed under three different anodic conditions: (i) broth in buffer (P-LB) (ii) broth with inoculum (P-LBG) and (iii) broth with inoculum and NR (P-LBG-NR) as mediator, in potential range of -0.6 to 0.8 V at 50 mVs-1.